본문 바로가기

Environmental controller

The HT-X1 system supports a stage-top incubator to maintain an optimal temperature and gas concentration conditions for live cell imaging. The incubator contains a built-in gas mixer unit and a chamber unit for mounting specimens onto the stage.

Environmental controller

Specifications

  • Model

    STXG-WSKMXA22B (Tokai HIT)

  • Dimensions

    151 (W) x 263 (D) x 196 (H) mm

  • Weight

    3.8 kg

  • Temperature setting range

    Sample temperature: 37 °C

    Top heater: 10 °C – 65 °C

    Bath heater: 10 °C – 50 °C

    Stage heater: 10 °C – 50 °C

    Lens heater: 10 °C – 45 °C

  • Rise time

    10 min to reach 50 °C

  • Temperature accuracy

    Within 0.3 °C

  • Humidity control

    Heated humidification by the heating bath unit

    Recommended water volume: 32 mL

  • CO₂ concentration range

    5% – 20%

    Control method: PID control

    Accuracy: ±0.1%

  • Gas type

    100% CO₂

  • Gas pressure

    Input: 0.1 MPa – 0.15 MPa

    Output: 160 mL per min

  • Power source

    100 - 240 V AC ±10%, 50/60 Hz

  • Maximum power consumption

    110 W

Ordering information

  • Cat. No. 900101-01

    Environmental Controller